fields separated by commas.
Query Syntax: *IDN?
Returned Parameters: Company Name, Power Supply Model, S/N (string format)
Example: Magna-Power Electronics, Inc., XR16-375, S/N: 1162-0361, F/W:1.0
Related Commands: CALibrate:IDN (Refer to Section 4.3.9.1, Calibrate Subsystem, for
information on configuring the XR Series power supply.)
4.5.7 Save
This command stores the present state of the power supply at the specified location in memory.
Data can be saved in 100 memory locations. (For XR Series power supplies, only memory
location 0 is available.) The following power supply parameters are stored by *SAV:
VOLT[:LEV][:IMM]
CURR[:LEV][:IMM]
VOLT:PROT[:LEV]
CURR:PROT[:LEV]
PER
Command Syntax: *SAV <NR1>
Parameters: 0 - 99
Example: *SAV 1
Query Syntax: None
Related Commands; *RCL, MEM
4.5.8 Recall
This command restores the power supply to a state that was previously stored in memory location
with a *SAV command. The *RCL command can recall data in 100 memory locations. For XR
Series power supplies, only memory location 0 is available. The following parameters are
recalled:
VOLT[:LEV][:IMM]
CURR[:LEV][:IMM]
VOLT:PROT[:LEV]
CURR:PROT[:LEV]
PER
79